Automatic Western Blotter XD236

Automatic Western Blotter XD236

Short Description:

The automatic western blotting instrument is an intelligent automatic operation protein (immune) blotting (Western blot) instrument, the product is suitable for the automatic operation of western blotting or other similar method kits.

-No guard is required when running

-Chinese and English man-machine dialogue

-Can detect 36 bands at the same time

-New high and low double liquid level oblique mouth negative pressure suction nozzle to ensure that the instrument is free of crossover and pollution.

-Fully enclosed piping system, with safety guarantee for operators.

-Can lock the pre-programmed program and unlock the pre-programmed program.

-Prompt alarm function.

Clinical application

-Mainly used for protein (Immuno) blotting or other similar methods for automated operation of kits.

-Autoimmune diseases: ENA, ANA, autoimmune liver disease, diabetes autoantibodies

-Viral infection: HIV, HCV, HEV, human T lymphocyte virus, etc.

-Bacteria and parasitic infections: Helicobacter pylori (HP), spirochetes, syphilis, toxoplasma, etc.

-Specific allergens


Product Detail

Product Tags

Specification

Test strip capacity

1-36 or 1-30 strips

programming capacity

≥ 38+2(cured a program)

liquid distribution system of liquid distribution system

6 channel

distribution range of liquid pump

0000μL – 3000μL / 1μL Increment 0000μL - 3000μL / 1μL

quantitative correction

Semi automatic quantitative correction

Incubation time / increment

0 h – 24 h / 1 min recyclable

The incubation time display

hours minutes and seconds (hour min second)

Rocking speed

3 gear optional

Communication

RS232 interface specification baud rate 9600 data bit 8 stop bit 1 parity check NO

The capacity of the reagent bottle

200 mL × 5

The volume of the washing bottle

1000 mL × 1

The volume of the waste liquid bottle

2000 mL× 1(no leaky closed negative pressure waste liquid bottle)

Power supply

AC 220V ±22V ;50Hz±1 Hz

Rated power

< 250VA

accuracy of the reagent distribution of the liquid dispensing pump

+ 3% (when the distribution of the reagents of each channel is 2000 mu L)

Precision of the liquid pump reagent dispensing

CV ≤5%(when the channel allocation amount of reagent is 2000 L)

The temperature control of the rocking bed

The temperature at room temperature and constant temperature 37℃、30℃、25℃

Temperature control accuracy

± 1..5℃

Cross contamination

When negative and positive samples are tested in alternating order, negative samples should not be presented with positive strip operation. According to the requirements of protein imprinting kit and testing procedure set.

Pre test

reagents preparation and pipeline cleaning before use

Test

add sample by hand

Results interpretation

According to the requirements of the kit and / or instrument operation instructions, the standard strip, manual or instrumental interpretation